Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K243400

    Validate with FDA (Live)

    Date Cleared
    2025-04-25

    (176 days)

    Product Code
    Regulation Number
    866.3981
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The cobas liat SARS-CoV-2 & Influenza A/B v2 nucleic acid test is an automated rapid multiplex real-time reverse transcription polymerase chain reaction (RT-PCR) test intended for the simultaneous qualitative detection and differentiation of severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2), influenza A virus and influenza B virus nucleic acids in anterior nasal (nasal) and nasopharyngeal swab specimens from individuals exhibiting signs and symptoms of respiratory tract infection. Clinical signs and symptoms of respiratory tract infection due to SARS-CoV-2 and influenza can be similar. This test is intended to aid in the differential diagnosis of SARS-CoV-2, influenza A and influenza B infections in humans and is not intended to detect influenza C virus infections.

    Nucleic acids from the viral organisms identified by this test are generally detectable in nasopharyngeal and nasal swab specimens during the acute phase of infection. The detection and identification of specific viral nucleic acids from individuals exhibiting signs and symptoms of respiratory tract infection are indicative of the presence of the identified virus, and aid in diagnosis if used in conjunction with other clinical and epidemiological information and laboratory findings.

    The results of this test should not be used as the sole basis for diagnosis, treatment, or other patient management decisions. Positive results do not rule out coinfection with other organisms. The organism(s) detected by the cobas liat SARS-CoV-2 & Influenza A/B v2 nucleic acid test may not be the definite cause of disease. Negative results do not preclude SARS-CoV-2, influenza A virus or influenza B virus infections.

    Device Description

    The cobas liat SARS-CoV-2 & Influenza A/B v2 nucleic acid test is performed on the cobas liat analyzer which automates and integrates sample purification, nucleic acid amplification, and detection of the target sequence in biological samples using real-time PCR assays. The assay targets both the ORF1 a/b non-structural region and membrane protein gene that are unique to SARS-CoV-2, a well-conserved region of the matrix gene of influenza A (Flu A target), and the nonstructural protein 1 (NS1) gene of influenza B (Flu B target). An Internal Control (IC) is included to control for adequate processing of the target virus through all steps of the assay process and to monitor the presence of inhibitors in the RT-PCR processes.

    AI/ML Overview

    This document describes the validation study for the cobas liat SARS-CoV-2 & Influenza A/B v2 nucleic acid test.

    Here's an analysis of the acceptance criteria and the study proving the device meets them:

    1. Table of Acceptance Criteria and Reported Device Performance

    Since this is a diagnostic test, the primary acceptance criteria revolve around analytical and clinical performance metrics like Limit of Detection, Inclusivity, Cross-Reactivity, Reproducibility, and Clinical Agreement (Positive Percent Agreement and Negative Percent Agreement). The document doesn't explicitly state "acceptance criteria" values in a separate table, but these are implied by the performance metrics reported and the general standards for diagnostic device clearance. I will extract the reported device performance from the provided text.

    Performance MetricTarget AnalyteSpecimen TypeReported Performance (Value)Implied Acceptance Criteria (Typically high for diagnostic tests)
    Analytical Sensitivity (LoD)SARS-CoV-2Co-spiked panels0.0350 TCID50/mLLowest detectable concentration for 95% positivity
    Influenza ACo-spiked panels0.00325 TCID50/mLLowest detectable concentration for 95% positivity
    Influenza BCo-spiked panels0.183 TCID50/mLLowest detectable concentration for 95% positivity
    Reactivity/InclusivitySARS-CoV-2Respective variants100% detection at 3x LoDDetection of various strains/variants
    Influenza ARespective variants100% detection at varying LoD (up to 12x)Detection of various strains/variants
    Influenza BRespective variants100% detection at 3x LoDDetection of various strains/variants
    Cross-Reactivity/Microbial InterferenceAll targetsVarious microorganismsNo cross-reactivity/interferenceNo false positives or interference from other common pathogens
    Competitive InhibitionAll targetsCo-spiked samplesNo interferenceAccurate detection of all targets even in co-infection
    Endogenous/Exogenous InterferenceAll targetsVarious substancesNo interferenceRobust performance in presence of common respiratory interferents
    Reproducibility (Negative)N/ANegative samples100.0% AgreementHigh agreement for negative samples across sites, lots, days
    Reproducibility (1x-2x LoD)SARS-CoV-2Low Positive samples100.0% AgreementHigh agreement for low positive samples
    Influenza ALow Positive samples99.6% AgreementHigh agreement for low positive samples
    Influenza BLow Positive samples99.6% AgreementHigh agreement for low positive samples
    Reproducibility (3x-5x LoD)SARS-CoV-2Moderate Positive100.0% AgreementHigh agreement for moderate positive samples
    Influenza AModerate Positive100.0% AgreementHigh agreement for moderate positive samples
    Influenza BModerate Positive100.0% AgreementHigh agreement for moderate positive samples
    *Clinical Performance (PPA)ProspectiveSARS-CoV-2NPS94.5% (90.7-96.8 CI)High sensitivity (ability to detect true positives)
    SARS-CoV-2ANS96.7% (93.4-98.4 CI)High sensitivity (ability to detect true positives)
    Influenza ANPS100.0% (93.4-100.0 CI)High sensitivity (ability to detect true positives)
    Influenza AANS100.0% (93.2-100.0 CI)High sensitivity (ability to detect true positives)
    Influenza BNPS100.0% (85.1-100.0 CI)High sensitivity (ability to detect true positives)
    Influenza BANS100.0% (86.2-100.0 CI)High sensitivity (ability to detect true positives)
    *Clinical Performance (NPA)ProspectiveSARS-CoV-2NPS97.6% (96.7-98.3 CI)High specificity (ability to correctly identify true negatives)
    SARS-CoV-2ANS97.2% (96.2-97.9 CI)High specificity (ability to correctly identify true negatives)
    Influenza ANPS99.3% (98.8-99.6 CI)High specificity (ability to correctly identify true negatives)
    Influenza AANS99.3% (98.8-99.6 CI)High specificity (ability to correctly identify true negatives)
    Influenza BNPS99.3% (98.8-99.6 CI)High specificity (ability to correctly identify true negatives)
    Influenza BANS99.5% (99.0-99.7 CI)High specificity (ability to correctly identify true negatives)
    *Clinical Performance (PPA)RetrospectiveInfluenza BNPS100.0% (89.8-100.0 CI)High sensitivity (ability to detect true positives)
    Influenza BANS100.0% (89.8-100.0 CI)High sensitivity (ability to detect true positives)
    *Clinical Performance (NPA)RetrospectiveInfluenza BNPS97.9% (94.7-99.2 CI)High specificity (ability to correctly identify true negatives)
    Influenza BANS98.3% (95.0-99.4 CI)High specificity (ability to correctly identify true negatives)

    2. Sample Sizes Used for the Test Set and Data Provenance

    • Prospective Clinical Study:

      • Sample Size: 1729 symptomatic subjects enrolled.
        • 1705 evaluable NPS specimens for analysis (19 non-evaluable due to missing/invalid results, 5 due to handling).
        • 1706 evaluable ANS specimens for SARS-CoV-2 and Influenza B analysis (22 non-evaluable due to missing/invalid results, 1 due to handling).
        • 1704 evaluable ANS specimens for Influenza A analysis (2 additional found inconclusive for comparator).
      • Data Provenance: Prospective, collected between September 2023 and March 2024 at 14 point-of-care testing sites in the United States (US).
    • Retrospective Clinical Study (Influenza B Supplement):

      • Sample Size: 223 archived NPS specimens and 206 archived ANS specimens (total 429).
        • One NPS sample pre-characterized as positive for influenza B was non-evaluable.
      • Data Provenance: Retrospective, frozen archived (Category III) specimens collected between 2019 and 2023. Distributed to 6 sites for testing.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Experts

    The document does not mention the use of experts to establish ground truth for the clinical test sets. For molecular diagnostic tests like this, the ground truth is typically established by comparing the investigational device's results against a highly accurate, accepted comparator method (another FDA-cleared Nucleic Acid Amplification Test specific for the target analytes). The expertise lies in the development and validation of these comparator methods, not in individual expert review of each sample for ground truth in this context.

    4. Adjudication Method for the Test Set

    The document describes discrepant result analysis for both prospective and retrospective clinical studies.

    • For the prospective study, "discrepant NAAT results" are detailed for SARS-CoV-2 (NPS and ANS), Influenza A (NPS and ANS), and Influenza B (NPS and ANS).
    • For the retrospective study, discrepant NAAT results are detailed for Influenza B (NPS and ANS).

    The method appears to be:

    • The cobas liat test result is compared to the FDA-cleared comparator NAAT result.
    • When there's a discrepancy (e.g., cobas liat positive, comparator negative), it explicitly states how many were "positive" and "negative" upon further investigation or re-evaluation (e.g., with "discrepant NAAT results").
      • For example: "Of 12 specimens negative on cobas® liat and positive on the comparator, 8 were positive and 4 were negative." This implies some form of re-testing or deeper analysis (not specified as "adjudication by experts" but rather "discrepant NAAT results"). It's more of a re-confirmation of the comparator or a third method, rather than a human expert consensus process. Such re-evaluation often involves re-testing using the comparator or a reference method.

    Therefore, while there's no "2+1" or "3+1" expert adjudication method described as would be seen in imaging studies, there is a discrepant resolution process based on further NAAT results. It's not "none" in the sense that discrepancies are just reported without follow-up; rather, they are further investigated using additional NAAT results to re-confirm the original comparator status if possible.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    No. This is a standalone diagnostic test (RT-PCR), not an AI-assisted imaging device or a test that involves human "readers" interpreting results. Therefore, an MRMC comparative effectiveness study involving human readers and AI assistance is not applicable and was not performed.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Yes, implicitly. This is a fully automated RT-PCR test run on the cobas liat analyzer. The performance metrics (LoD, inclusivity, cross-reactivity, reproducibility, and clinical agreement) are measures of the device's performance on its own against established ground truth (comparator NAAT). While humans load samples and interpret the final digital result (positive/negative), the core detection and differentiation is algorithm-driven within the instrument, making its performance essentially "standalone" in the context of diagnostic accuracy.

    7. The Type of Ground Truth Used

    • Clinical Performance (Prospective and Retrospective): The ground truth for clinical sample testing was established by comparing the cobas liat results against an FDA-cleared Nucleic Acid Amplification Test (NAAT), which serves as the reference or "ground truth" method for molecular diagnostic assays. The document explicitly states: "PPA and NPA were determined by comparing the results of cobas® liat SARS-CoV-2 & Influenza A/B v2 to the results of an FDA-cleared Nucleic Acid Amplification Test (NAAT)." and "The comparator method was an acceptable FDA-cleared molecular assay."
    • Analytical Studies (LoD, Inclusivity, Cross-Reactivity, Interference, Reproducibility): Ground truth was established by preparing precisely known concentrations of viral material (cultured or inactivated viruses) or specific microorganisms in controlled laboratory settings. For these studies, the "ground truth" is meticulously prepared and verified laboratory standards.

    8. The Sample Size for the Training Set

    The document does not specify a separate "training set" sample size. For an RT-PCR diagnostic platform, the "training" involves the fundamental biochemical and optical engineering, and the optimization of assay (reagent) design to achieve sensitivity and specificity. This is distinct from machine learning models that often require large, labeled datasets for "training." The analytical and clinical validation studies described here are verification and validation (V&V) studies, akin to a "test set" to prove the device's performance against its design specifications and clinical utility.

    9. How the Ground Truth for the Training Set Was Established

    Since no explicit "training set" for a machine learning algorithm is mentioned (as this is a molecular diagnostic test), this question is not directly applicable. However, the ground truth for assay development and optimization (which can be considered analogous to "training" in a broader sense of device development) would have been established through extensive laboratory work using:

    • Highly characterized viral cultures or purified nucleic acids: Used to define target sequences, optimize primer/probe design, and determine initial analytical sensitivity.
    • Spiked samples: Adding known quantities of targets or interferents to negative clinical matrices to mimic real-world conditions during early development.
    • Early clinical samples: Used to refine assay performance and resolve initial issues prior to formal validation studies.

    These processes ensure the assay correctly identifies the target nucleic acids.

    Ask a Question

    Ask a specific question about this device

    K Number
    K243346

    Validate with FDA (Live)

    Date Cleared
    2025-04-11

    (165 days)

    Product Code
    Regulation Number
    866.3982
    Age Range
    0 - 18
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The cobas® liat SARS-CoV-2 v2 nucleic acid test is an automated real-time reverse transcription polymerase chain reaction (RT-PCR) test intended for the qualitative detection of severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) nucleic acids in anterior nasal (nasal) and nasopharyngeal swab specimens collected from individuals exhibiting signs and symptoms of respiratory tract infection (i.e., symptomatic). Additionally, this test is intended to be used with nasal and nasopharyngeal swab specimens collected from individuals without signs and symptoms of COVID-19 (i.e., asymptomatic).

    The cobas® liat SARS-CoV-2 v2 nucleic acid test is intended for use as an aid in the diagnosis of COVID-19 if used in conjunction with other clinical and epidemiological information and laboratory findings. SARS-CoV-2 RNA is generally detectable in nasal swab and nasopharyngeal swab specimens during the acute phase of infection.

    Positive results are indicative of the presence of SARS-CoV-2 RNA. Positive results do not rule out co-infection with other microorganisms. Negative results do not preclude SARS-CoV-2 infection. Negative results must be combined with clinical observations, patient history, and epidemiological information. The results of this test should not be used as the sole basis for diagnosis, treatment, or other patient management decisions.

    A negative result from an asymptomatic individual is presumptive. Additionally, a negative result obtained with a nasal or nasopharyngeal swab collected from an asymptomatic individual should be followed up by testing at least twice over three days with at least 48 hours between tests.

    Device Description

    The cobas® liat SARS-CoV-2 v2 nucleic acid test is performed on the cobas® liat analyzer which automates and integrates sample purification, nucleic acid amplification, and detection of the target sequence in biological samples using real-time PCR assays. The assay targets both the ORF1 a/b non-structural region and membrane protein gene that are unique to SARS-CoV-2. An Internal Control (IC) is included to control for adequate processing of the target virus through all steps of the assay process and to monitor the presence of inhibitors in the RT-PCR processes.

    AI/ML Overview

    The provided text is a 510(k) Clearance Letter for a medical device which does not include information about AI/ML models. Therefore, it's not possible to extract the information you requested about Acceptance Criteria and a study proving an AI/ML device meets those criteria.

    The device described, the "cobas liat SARS-CoV-2 v2 nucleic acid test," is an in vitro diagnostic (IVD) device based on real-time RT-PCR technology. It directly detects viral targets and its performance is evaluated through analytical and clinical studies common for IVDs, not against AI/ML performance metrics like sensitivity, specificity, MRMC studies, or multi-reader reviews.

    Here's a breakdown of why your requested information isn't present in this document:

    • No AI/ML Component: The document describes a traditional RT-PCR assay. There is no mention of algorithms, machine learning, deep learning, or any AI component.
    • Performance Metrics Differ: The performance metrics provided (Limit of Detection, Inclusivity, Cross-reactivity, Reproducibility, Positive Percent Agreement, Negative Percent Agreement) are standard for IVD assays. They are not analogous to metrics used for evaluating AI/ML models (e.g., AUC, F1-score, accuracy in image classification, or diagnostic improvement from AI-assistance).
    • No Human Reader Interaction: Since it's an automated lab test, there's no "human reader" (like a radiologist) involved in interpreting the device's output in a way that an AI would assist. The output is qualitative (Detected/Not Detected).
    • No Ground Truth Experts in the AI Sense: Ground truth for this IVD is established by a "comparator" (another FDA-cleared NAAT) and clinical/epidemiological information, not by multiple human experts reviewing AI outputs or images.
    • No Training/Test Set Split for AI: The "test set" and "training set" concepts described in your request are fundamental to AI/ML model development and validation. For this IVD, there's a "clinical performance evaluation" using prospective and retrospective samples, which serves as the validation dataset, but it's not structured as a training/test split for an AI.

    Therefore, I cannot provide the requested table and detailed points because the provided document does not pertain to an AI/ML device.

    If you have a document describing an AI/ML medical device, please provide that, and I can attempt to extract the relevant information.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1